Comparision of Several Preprocessing Algorithms based on Near Infrared Spectroscopic Measurement of Glucose in Aqueous Glucose Solutions
Abstract
Glucose concentration measurement is the basis of noninvasive detection of blood glucose concentration. It is significant in scientific research. In this study, Near Infrared Spectrum (NIRS) and regression analysis methodology were combined to measure the glucose concentration. The spectrum of glucose solutions was obtained with the Fourier Transformed Infrared Spectrometer, and then the data was used for regression analysis. In addition, the method of Paritial Least Squares (PLS) was used to achieve principle components and various spectral preprocessing methods were discussed. During PLS modeling, the Savitzky-Golay could improve the Prediction Residual Error Sum of Squares (PRESS) within 6%. The experiment results demonstrate that NIRS has the potential for the measurement of glucose solution.
Full Text:
PDFRefbacks
- There are currently no refbacks.
This work is licensed under a Creative Commons Attribution-ShareAlike 4.0 International License.
Indonesian Journal of Electrical Engineering and Computer Science (IJEECS)
p-ISSN: 2502-4752, e-ISSN: 2502-4760
This journal is published by the Institute of Advanced Engineering and Science (IAES) in collaboration with Intelektual Pustaka Media Utama (IPMU).