A Curve-fitting Calibration Method applied for Ultrasonic Flow-meter
Abstract
As the influence of fluid distribution in the internal pipe, the measurement characteristics of theory and practice exist significant differences in Ultrasonic Flow-meter(USF). Through analysis of fluid state, the method of curve-fitting is applied for the calibration of USF. Experimental results show that the USF can achieve level-1 accuracy with just a correction of 5 flow points, and this method performs a low computational complexity and strong practicality.
DOI: http://dx.doi.org/10.11591/telkomnika.v11i10.3392
Keywords
Full Text:
PDFRefbacks
- There are currently no refbacks.
This work is licensed under a Creative Commons Attribution-ShareAlike 4.0 International License.
Indonesian Journal of Electrical Engineering and Computer Science (IJEECS)
p-ISSN: 2502-4752, e-ISSN: 2502-4760
This journal is published by the Institute of Advanced Engineering and Science (IAES) in collaboration with Intelektual Pustaka Media Utama (IPMU).